We offer onsite MFC calibration of Alicat and Bronkhorst MFC from 20nmlpm - 8000 nlpm flow range.

MFCCal

MFCs will be calibrated in their installed
position. MFCs will be calibrated with
customer gas infrastructure, including its
limitations. Available for H2, N2, O2, Air and
CO2 (O2 and CO2 with Bronkhorst
correction factors). With TIRRON
calibration certificates. Does not include repair of MFCs, which cannot be calibrated.
(MFC can not be calibrated within our tolerances) TIRRON Reference is
calibrated at ISO 17025 certified laboratory with 1 year calibration interval. With 5
calibration points: 0, 20, 40, 60, 80 and 100
% FS. Available ranges 20 nmlpm - 8000
nlpm.
Tolerance and Uncertainty:
20nmlpm - 1000 nlpm:
-Tolerance: 1 % Measurement + 0.2 % Full
Scale of DUT
-Uncertainty: 0.5 % Measurement + 0.1%
Full Scale of used References